Uploaded image for project: 'Data Management'
  1. Data Management
  2. DM-29793

Camera Hexapod: The EUI needs to be restarted after re-plugging the EtherCat cable

    XMLWordPrintable

    Details

      Description

      After unplugging and re-plugging the EtherCAT cable:

      Step
      Unplug the EtherCAT cable between the control PC and the first Copley XE2 drive.

      Expected Result
      An EtherCAT Lost event is created and the system transitions to Fault state.

      Actual Result
      The EUI switched off without warning.

       

        Attachments

          Activity

            People

            Assignee:
            ttsai Te-Wei Tsai
            Reporter:
            hdrass Holger Drass
            Watchers:
            Holger Drass, Te-Wei Tsai
            Votes:
            0 Vote for this issue
            Watchers:
            2 Start watching this issue

              Dates

              Created:
              Updated:

                Jenkins